Accuracy Comparison Between Burettes and Pipettes: Fact-Checked & Updated 2025 Edition
Laboratory professionals frequently compare burettes and pipettes to determine which delivers greater accuracy and reliability for volumetric measurements. Comprehensive analysis below draws directly on current international standards (ISO 385:2005, ISO 648:2008, ASTM E287-02(2019), ISO 8655-2), manufacturer data, and best-practice recommendations, ensuring rigorous fact-checking for trustworthy publication.
Burette Design, Calibration & Standards
Structure:
Burettes are graduated tubes (mostly glass, sometimes plastic) fitted with a precision stopcock that enables controlled delivery. Traditionally used in titrations, burettes provide flexible, variable-volume dispensing from zero to their total capacity.
Types & Accuracy Classes:
ISO 385:2005 and ASTM E287-02 specify two primary accuracy classes:
Class A: High-precision; suitable for quantitative analysis and research; certified and individually serialized.
Class B: Standard precision; tolerances exactly double those of Class A; common in educational and routine settings.
Tolerance Table (ISO 385:2005/ASTM E287):
Volume (mL)
Subdivision (mL)
Class A Error (±mL)
Class B Error (±mL)
1
0.01
0.006
0.010
2
0.01
0.01
0.02
5
0.01–0.02
0.01
0.02
10
0.02–0.05
0.02/0.03
0.05
25
0.05–0.10
0.03/0.05
0.05/0.10
50
0.10
0.05
0.10
100
0.20
0.10
0.20
Source: ISO 385:2005 Table 1, ASTM E287-02, DIN EN ISO 385, Manufacturer guides
Reference Temperature: All tolerances are valid at 20°C unless specified for tropical regions where 27°C is allowed by ISO 384.
Reading & Use: Skilled users regularly interpolate burette readings to 0.05 mL or better. Always rinse with target solution prior to titration to avoid dilution errors.
Pipette Design, Calibration & Standards
Types & Uses:
Volumetric (Bulb) Pipettes: Deliver one precise volume; highest accuracy.
Graduated Pipettes: Deliver a range of volumes; less accurate than volumetric.
Micropipettes: Fixed or adjustable; suited for sub-milliliter measurements in molecular biology.
Calibration:
TD (To Deliver / Ex): Dispensed volume corresponds to calibration; residual left in tip is accounted for and not expelled.
TC (To Contain / In): Holds the volume specified; must be blown out to deliver the full calibrated amount.
Tolerance Table (ISO 648:2008 and Manufacturer Data):
Volume (mL)
Class A Error (±mL)
Class B Error (±mL)
0.5
0.005
0.010
1
0.008
0.015
2
0.01
0.02
5
0.015
0.03
10
0.02
0.04
20
0.03
0.06
25
0.03
0.06
50
0.05
0.10
100
0.08
0.15
Source: ISO 648:2008 Table 1, DIN ISO 648, manufacturer guides
Comparative Analysis
Feature
Burette
Pipette
Calibration
TD (Deliver)
Usually TD; TC possible
Accuracy Range
±0.006–0.10 mL (Class A)
±0.005–0.08 mL (Class A)
Best Use
Variable delivery, titration steps
Fixed-volume transfer, aliquoting
Error Sources
Meniscus reading, stopcock leaks
Meniscus, handling, tip residue
Technique Required
Eye-level parallax; interpolate
Maintain vertical, avoid air, no blowout (except TC)
Ready to Master Lab Precision?
Take this quick 8-question quiz to test your knowledge on burette vs pipette accuracy - essential for US college chemistry, biotech labs, and DIY science experiments. Finish to get your score, personalized US Prime recommendations from Amazon (1-2 day delivery nationwide).
Time left:
What is the typical Class A tolerance for a 10 mL volumetric pipette versus a 50 mL burette?
Meniscus Reading: Always at eye level to avoid parallax errors in both burettes and pipettes.
Burette Handling: Remove air bubbles in jet and stopcock before titrations. Use slow, controlled delivery around endpoint for highest accuracy.
Pipette Handling: Pre-wet tip, avoid bubbles, do not blow out TD pipettes (unless TC type). Pipette angle should remain close to vertical.
Technological Updates (2025)
Smart Pipette: New devices with haptic feedback and digital logging have improved novice pipetting accuracy by ~30%, according to recent comparative studies.
ISO 8655-2: Pipette calibration now treats the pipette and tip as a system; changing tip requires recalibration.
Recommendations for Laboratory Practice
Use pipettes for fixed-volume, high-precision transfers (aliquots, standards for analytical procedures).
Use burettes for variable increments in titrations or when volume must be adjusted during an experiment.
Always check your instrument’s calibration mark (TD or TC) and adjust technique accordingly.
Schedule routine calibration based on manufacturer frequency, usage, and any regulatory standards in force.
Employ emerging technologies for pipette handling and documentation where available.
Tables for Reference
ISO/ASTM Tolerance Table: Burettes vs. Pipettes
Instrument
Volume (mL)
Class A Error (±mL)
Class B Error (±mL)
Volumetric/Bulb Pipette
10
0.02
0.04
Volumetric/Bulb Pipette
25
0.03
0.06
Burette
25
0.03
0.05
Burette
50
0.05
0.10
Calibration Types
Type
Mark
Usage
To Deliver
TD / Ex
Do not blow out remaining liquid; residual accounted for
To Contain
TC / In
Blow out remaining liquid to deliver full volume
Conclusion
Burettes offer excellent flexibility for titration steps with robust standards for accuracy. Pipettes (particularly volumetric, Class A) provide unmatched precision for fixed-volume transfers. Selection should always be based on the demands of the experimental protocol—precision for fixed volumes favors pipettes, flexibility for variable volumes favors burettes. Instrument class, technical standards, and calibration protocols directly govern reliability, so careful documentation and routine maintenance are required for quality laboratory practice.